What is the principle of Laura system?
Laura works on Reflectance photometry
Is Laura is an open system?
It is not open system. It is compatible for Erba Lachema strips only.
Is Laura a replacement for Visual Microscopy?
Sediments such as crystals and casts and the type of bacterial cells are possible to be seen in visual microscopy. Hence neither Laura Smart nor any Urine analyzer can replace Visual microscopy. Laura Smart is meant for routine screening of urine samples.
What is major difference between LAURA SMART & LAURA new version?
LAURA SMART is single strip reader whereas LAURA new version is continuous loading system for higher throughput of 400 tests per hour.
Does this system require daily calibration by user?
Laura does not require daily calibration.
What is the inbuilt memory capacity of Laura ?
Memory capacity of Laura is 2000 results. After 2000 results the 1st result will automatically deleted.
What other maintenance is required daily for Laura analyser?
Cleaning of attached waste bin & conveyor belt is required.
Does Laura automatically read different strips from Eba Lachema?
Yes, it automatically reads the strips without changing any settings.
Does Laura automatically calculate albumin creatinine ratio?
Yes it calculates the ACR ratio automatically by using special algorithms.