Chem 5x
Semi Automated Clinical Chemistry Analyzer
![](https://erba.velocitabrandconsultants.com/wp-content/uploads/2024/05/chem5x-1.jpg)
Chem 5x
Semi Automated Clinical Chemistry Analyzer
Meet the Chem 5x
Product Specifications
![1 1](https://erba.velocitabrandconsultants.com/wp-content/uploads/elementor/thumbs/1-qofdqwcnzomyyvdvvbs7x64oo13ppl213mdfp6lwn4.png)
Photometer
Static Photometer with 6 IAD filters + 2 optional filters
Your Benefit:
Ensures precision for longer time with zero fungal growth
![](https://erba.velocitabrandconsultants.com/wp-content/uploads/2024/03/psf4.png)
Cuvette System
Dual cuvette system
Your Benefit:
Ensures flow cell safety and offers flexibility to perform immunoturbidimetry with precision
![4 4](https://erba.velocitabrandconsultants.com/wp-content/uploads/elementor/thumbs/4-1-qoh5a32xszmhgp4aaz5i308l2rulfr12w58wprdg8w.png)
Program Memory
250 user defined test programs
Your Benefit:
Covers the entire panel of chemistry testing
![Picture5 Picture5](https://erba.velocitabrandconsultants.com/wp-content/uploads/elementor/thumbs/Picture5-qofvd24qvsr7vy8ctg2y5544mjemu6moafajpqbyps.png)
Result Memory
Archival of upto 10,000 test results
Your Benefit:
Helps in managing patient records effectively
![pic5 pic5](https://erba.velocitabrandconsultants.com/wp-content/uploads/elementor/thumbs/pic5-qog0n8t5p13szlz69nlgn5dnl6089hnifr7bd8cbr4.png)
Assay Modes
9 assay modes
Your Benefit:
Best testing practice to perform various chemistry tests with dedicated modes.
![f4 f4](https://erba.velocitabrandconsultants.com/wp-content/uploads/elementor/thumbs/f4-qofj60okolqr0tyni6380fkbwhdojoo4xdkz1zav0g.png)
Lamp Saving Feature
Auto Lamp life saving feature
Your Benefit:
Better life span and optimise usage of the lamp
End Point Reactions: Glucose, Cholesterol, Triglycerides, Albumin, Total Protein, Uric Acid, Calcium, Magnesium, Bilirubin Total, Bilirubin Direct
Kinetic Reactions: SGOT, SGPT, GGT, LDH, CK, CK MB, ALP, ACP, Amylase, Lipase
Two Point Rate Reactions: Urea, Creatinine
Immunoturbidimetric Chemistries: RA, CRP, ASO, Direct HDL, Direct LDL, Apo A, ApoB, Lp (a)
Special Chemistries: HbA1c, Iron, Ferritin, Phosphorus, Microprotein, Ammonia, Bicarbonates
System General
System Type
Discrete, automated, random access, patient prioritized, clinical chemistry analyzer, 1,2, 3 reagent system
Throughput
Upto 1040 tests/hr (800 tests/hr) photometric and 240 tests/hr with ISE (4 Channel)*
Analytical Methods
1-Point, 2-Point, Rate-A, Rate-B, Direct Potentiometry
Barcode Reader
For reagents and samples
Reaction Tray
147 hard glass cuvettes
Reaction Mixing
Stirrer with variable speed
Reading Volume
Reaction volume 150 ul
On-board Laundry
7 stage cleaning, 2 stage drying
Photometer
Multi-wavelength diffraction grating with 15 wavelengths (340, 376, 405, 450, 480, 505, 528, 546, 570, 600, 628, 660, 700, 750 and 800 nm)
Calibration
K-Factor, Linear (1, 2 point & multipoint), 4P and SP Logit-log, cubic spline, exponential, polynomial. On-board serial dilution for calibrator
OD Range
0.0 - 3.0
Light Source
Tungsten halogen lamp
Detector
Silicon photo-diode
Water Consumption
<25 litres /hr
Programmable Parameters:
Default system pack parameters + upto 99 user defined parameters, Unlimited profiles and unlimited calculation items
Quality Control
QC plot data with QC rules. Provision for lab mean. Twin plot
Power Requirement
AC220V± 10%,50 Hz or AC 110 V± 10%, 60 Hz
Power Consumption
2000VA.
Dimensions (mm)
Approx . 1350 (L) x 1000 (D) x 1250 (H)
Weight
Approx . 275 kgs.
Sample Handling
Sample
Serum, Urine, Plasma, CSF, whole blood, other.
Standard / STAT Carousal
44 positions
Sample Rack
Upto 150 samples can be placed simultaneously (excluding standards, calibrators & controls)
Sample Pipetting
2 -60 µl (adjustable in 0.1 µIstep) for Biochemistry, 70 µl (fixed) for ISE Capacitance probe with level sensing, vertical obstruction detection, Clot detection and serum indices
Auto Rerun
Repeat with same, increased or decreased volume (upto 1:150)
Sample Tubes/ Cups
Primary tubes of 5 ml, 7 ml, 10 ml and sample cups
Reagent Handling
Reagent Tray
86 reagent positions with on-board refrigeration
Reagent Pipetting
Two separate probes for Rl, R2 & R3 Rl:60-300 µl (adjustable in 0.5 µl)
R2
0, or 10- 200 µl (adjustable in 0.5 µl)
R3
0, or 10- 200 µl (adjustable in 0.5 µl) Capacitance probe with liquid level sensing & vertical obstruction detection
System Interface
Analyzer - PC
RS - 232 C Serial Port PC - Host
Computer
Bidirectional TCP/IP & RS-232
PC - Printer
USB
Operating system
Windows-based
Database
Unlimited results