EM 360
Fully Automated Clinical Chemistry Analyzer
EM 360
Fully Automated Clinical Chemistry Analyzer
Meet the EM 360
Product Specifications
Performance
Proven diffraction grating technology
offers uncompromised reliability in terms of
accuracy and precision of test results.
Capacitance probes provide liquid level
sensing & separate probes ensure
reliability of throughput
Your Benefit:
No data…
Enviorment Friendly
Permanent hard glass cuvettes
eliminate recurring cost of
disposable cuvettes and avoid
plastic usage.
Fully equipped with a multi stage
laundry station.
Your Benefit:
No data…
Reliability
Daily and monthly Levey-Jennings curve and facility to run user defined controls, ensures reliable reports.
Your Benefit:
No data…
Cloud-Based / Service
Integrated Internet of Things (IoT) sensors To enhance efficiency.
Your Benefit:
No data…
Flexibility
Emergency sample processing facility provides total flexibility for reporting the stat patient reports.
Your Benefit:
No data…
Flexibility
Offers connectivity to the host computer via LIS Automatically schedules daily workload, implements test instructions, Q.C. needs and monitors test run, verifies individual test results, repeats or
Your Benefit:
No data…
End Point Reactions: Glucose, Cholesterol, Triglycerides, Albumin, Total Protein, Uric Acid, Calcium, Magnesium, Bilirubin Total, Bilirubin Direct
Kinetic Reactions: SGOT, SGPT, GGT, LDH, CK, CK MB, ALP, ACP, Amylase, Lipase
Two Point Rate Reactions: Urea, Creatinine
Immunoturbidimetric Chemistries: RA, CRP, ASO, Direct HDL, Direct LDL, Apo A, ApoB, Lp (a)
Special Chemistries: HbA1c, Iron, Ferritin, Phosphorus, Microprotein, Ammonia, Bicarbonates
System General
System Type
Discrete, automated, random access, patient prioritized clinical chemistry analyzer
Throughput
360 tests / hour photometric and
600 tests / hour with ISE*
(*optional ISE with Na+, K+, Cl-, Li+)
Analytical Methods
1-Point, 2-Point, Rate-A, Rate-B,Direct Potentiometry (optional)
Barcode Reader
For reagents and samples (optional)
Reaction Tray
60 hard glass cuvettes
Reaction Mixing
Stirrer with variable speed
Minimum reaction Volume
150 μl with maximum 200 μl reading volume
On-board Laundry
5 stage cleaning, 2 stage drying
Photometer
Multi-wavelength diffraction grating with12 wavelengths (340, 376, 415, 450, 480,505, 546, 570, 600, 660, 700, 750 nm)
Calibration
K-Factor, Linear (1, 2 point & multipoint),4P and 5P Logit-log, cubic spline, exponential, polynomial,On-board serial dilution for calibrator
OD Range
0.0 - 3.0
Light Source
Halogen lamp
Detector
Silicon photo-diode
Water Consumption
< 10 litres
Programmable Parameters:
Default system pack parameters+ upto 99 user defined parameters, Unlimited profile and unlimited calculation items
Quality Control
QC plot data with QC rules.Provision for lab mean. Twin plot
Power Requirement
AC 220 V ± 10 %, 50 Hz orAC 110 V ± 10 %, 60 Hz
Power Consumption
800 VA
Dimensions (mm)
Approx. 675 (W) x 840 (D) x 1120 (H)
Weight
Approx. 150 kgs.
Sample Handling
Sample
Serum, Plasma, Urine, CSF, Whole blood, other
Sample Unit
82 positions for samples, blank , controls, calibrators, STAT sample
Sample Rack
Upto 150 samples can be placed simultaneously (excluding standards, calibrators & controls)
Sample Pipetting
2 - 70 µl (adjustable in 0.1 µl step) for Biochemistry, 70 µl fixed for ISE Capacitance probe with liquid level sensing & vertical obstruction detection,serum indices
Auto Rerun
Repeat with same, increased or decreased volume ( upto 1:150)
Sample Tubes/ Cups
Primary tubes of 5 ml, 7 ml, 10 ml and sample cups
Reagent Handling
Reagent Tray
50 positions for reagents with onboard cooling
Reagent Pipetting
Capacitance probe with level sensing & vertical obstruction detection
R1
R1: 50 - 300 µl (adjustable in 1 µl step)
R2
R2: 10 - 200 µl (adjustable in 1 µl step)
System Interface
Analyzer - PC
USB
Computer
Bidirectional TCP/IP & RS-232
PC - Printer
USB
Operating system
Windows-based
Database
Unlimited results