H 360
Automated 3 Part Differential Hematology Analyzer
![](https://erba.velocitabrandconsultants.com/wp-content/uploads/2024/05/1H360.jpg)
![](https://erba.velocitabrandconsultants.com/wp-content/uploads/2024/05/2H360.jpg)
H 360
Automated 3 Part Differential Hematology Analyzer
Meet the H 360
Product Specifications
![1 1](https://erba.velocitabrandconsultants.com/wp-content/uploads/elementor/thumbs/1-3-qoh5jwi96p2aq2uwz7x84j1wfnglty05kqjh5stf9c.png)
Parameters
Reports 24 parameters + 3 histograms
Your Benefit:
Multi-parameter estimation for easy and accurate reporting of all sample types
![2 2](https://erba.velocitabrandconsultants.com/wp-content/uploads/elementor/thumbs/2-3-qoh5jxg3dj3l1otjtqbup0td11bz1n3vwv6yn2s134.png)
Diffrentiator of IDA & Beta Thalassemia
Reports Mentzer index and RDWI
Your Benefit:
Enables differentiation between Iron deficiency Anaemia and Beta Thalassemia
![5 5](https://erba.velocitabrandconsultants.com/wp-content/uploads/elementor/thumbs/5-2-qoh5jydxkd4vdas6o8qh9iktmf7c9c7m8zug4cqmww.png)
Diluting tool in pediatric mode
ADDM (Auto Dispense Diluting Mechanism)
Your Benefit:
Auto dispenses predefined volume, reducing contamination and manual errors
![5 5](https://erba.velocitabrandconsultants.com/wp-content/uploads/elementor/thumbs/5-2-qoh5jydxkd4vdas6o8qh9iktmf7c9c7m8zug4cqmww.png)
Regents Inventory Management
RFID + RIMS for reagents
Your Benefit:
Helps in real time inventory management
![f5 f5](https://erba.velocitabrandconsultants.com/wp-content/uploads/elementor/thumbs/f5-14-qoh6kj9tj97l6znc2jysss9lv8wktafka1tph0i2rk.png)
Reporting Formats
Customized reporting formats
Your Benefit:
Saves morphological findings & other test results in single report
![6 6](https://erba.velocitabrandconsultants.com/wp-content/uploads/elementor/thumbs/6-2-qoh5jzbrr765owqtir53u0ca7t2ph1bcl4hxlmp8qo.png)
Storage
Storage: 50,000 Results with Histograms
Your Benefit:
More data storage allows an easy access to old patient samples for monitoring the clinical progress
![f7 f7](https://erba.velocitabrandconsultants.com/wp-content/uploads/elementor/thumbs/f7-8-qoh6kk7nq38villyx2dfda12gmry0zjam6h6yagolc.png)
Linearity
Linearity WBC - 0~300×10³/μL, PLT - 0 - 3000 x 10^9/L
Your Benefit:
Wide linearity in H360 helps to clinically analyze extreme patient demographics with leukemias, thrombocytosis, etc.
CBC Analysis
Technical Specification
Mode Type
Whole Blood, Predilute, Capillary
Total Parameters
24 Parameters including Mentzer index and RDWI, CBC: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PCT, PDW-CV, PDW-SD, P-LCR, P-LCC. 3 Part Diff: Lymph % & #, Mid % & #, Gran% & #
Principle of measurement
RBC/ PLT/ WBC/ DIFF: Electrical Impedance
Hb: Cyanide Free Colorimetry
MCV: Measured
HCT: Calculated
Histograms
3 (WBC/RBC/PLT)
Sample Volume
Whole Blood: 9 µL
Pre- diluted: 20 µL
Capillary: 20 µL
Linearity Range
WBC : 0 - 300 (x 103/µL)
RBC: 0.00 - 8.50 (x 106/µL)
Hb : 0 - 25.0 (g/dL)
HCT : 0 - 67 (%)
PLT : 0 - 3000 (x 103/µL)
Calibrator
ELite H Cal
Tri-level Controls
Elite H3 CON L, N, H
Open Vial Stability at 2-8˚C:
16 Days
Throughput
60 Tests/Hr
Sample Bar Code Reader
For reading sample bar codes
Quality Control
500 files with L-J graph
X-Bar Analysis
Data Storage
50,000 Results with Graphs
Dimension (mm)
364x477x417
Weight(Kg)
25
Reagents
Erba H 360 Dil
Erba H 360 Lyse
Elite H Clean
Interfaces
4 USB ports (For connecting external devices including external printer) + 1 LAN Port (HL 7, TCP/IP; LIS connectivity)
Operating Environment Temperature
15-40˚C
Humidity
30-90%
Power Requirement
A.C.100-240V with built-in voltage stabiliser 50/60Hz, 200VA
System General
System Type
Discrete, automated, random access, patient prioritized, clinical chemistry analyzer, 1,2, 3 reagent system
Throughput
Upto 1040 tests/hr (800 tests/hr) photometric and 240 tests/hr with ISE (4 Channel)*
Analytical Methods
1-Point, 2-Point, Rate-A, Rate-B, Direct Potentiometry
Barcode Reader
For reagents and samples
Reaction Tray
147 hard glass cuvettes
Reaction Mixing
Stirrer with variable speed
Reading Volume
Reaction volume 150 ul
On-board Laundry
7 stage cleaning, 2 stage drying
Photometer
Multi-wavelength diffraction grating with 15 wavelengths (340, 376, 405, 450, 480, 505, 528, 546, 570, 600, 628, 660, 700, 750 and 800 nm)
Calibration
K-Factor, Linear (1, 2 point & multipoint), 4P and SP Logit-log, cubic spline, exponential, polynomial. On-board serial dilution for calibrator
OD Range
0.0 - 3.0
Light Source
Tungsten halogen lamp
Detector
Silicon photo-diode
Water Consumption
<25 litres /hr
Programmable Parameters:
Default system pack parameters + upto 99 user defined parameters, Unlimited profiles and unlimited calculation items
Quality Control
QC plot data with QC rules. Provision for lab mean. Twin plot
Power Requirement
AC220V± 10%,50 Hz or AC 110 V± 10%, 60 Hz
Power Consumption
2000VA.
Dimensions (mm)
Approx . 1350 (L) x 1000 (D) x 1250 (H)
Weight
Approx . 275 kgs.
Sample Handling
Sample
Serum, Urine, Plasma, CSF, whole blood, other.
Standard / STAT Carousal
44 positions
Sample Rack
Upto 150 samples can be placed simultaneously (excluding standards, calibrators & controls)
Sample Pipetting
2 -60 µl (adjustable in 0.1 µIstep) for Biochemistry, 70 µl (fixed) for ISE Capacitance probe with level sensing, vertical obstruction detection, Clot detection and serum indices
Auto Rerun
Repeat with same, increased or decreased volume (upto 1:150)
Sample Tubes/ Cups
Primary tubes of 5 ml, 7 ml, 10 ml and sample cups
Reagent Handling
Reagent Tray
86 reagent positions with on-board refrigeration
Reagent Pipetting
Two separate probes for Rl, R2 & R3 Rl:60-300 µl (adjustable in 0.5 µl)
R2
0, or 10- 200 µl (adjustable in 0.5 µl)
R3
0, or 10- 200 µl (adjustable in 0.5 µl) Capacitance probe with liquid level sensing & vertical obstruction detection
System Interface
Analyzer - PC
RS - 232 C Serial Port PC - Host
Computer
Bidirectional TCP/IP & RS-232
PC - Printer
USB
Operating system
Windows-based
Database
Unlimited results