H 560
Automated 5 Part Differential Hematology Analyzer
![](https://erba.velocitabrandconsultants.com/wp-content/uploads/2024/05/1H560.jpg)
![](https://erba.velocitabrandconsultants.com/wp-content/uploads/2024/05/2H560.jpg)
![](https://erba.velocitabrandconsultants.com/wp-content/uploads/2024/05/3H560.jpg)
H 560
Automated 5 Part Differential Hematology Analyzer
Meet the H 560
Product Specifications
![1 1](https://erba.velocitabrandconsultants.com/wp-content/uploads/elementor/thumbs/1-3-qoh5jwi96p2aq2uwz7x84j1wfnglty05kqjh5stf9c.png)
Parameters
36 parameters
Your Benefit:
More clinical parameters reported, reduces smear review
![f2 f2](https://erba.velocitabrandconsultants.com/wp-content/uploads/elementor/thumbs/f2-14-qoh6kibzcf6avdop81k68ai59v17llbtxx67zqjgxs.png)
Diffrentiator of IDA & Beta Thalassemia
Reports Mentzer index and RDWI
Your Benefit:
Enables differentiation between Iron deficiency Anaemia and Beta Thalassemia
![f3 f3](https://erba.velocitabrandconsultants.com/wp-content/uploads/elementor/thumbs/f3-14-qoh6kibzcf6avdop81k68ai59v17llbtxx67zqjgxs.png)
Reporting
Reporting of NLR & NRBC
Your Benefit:
- NLR a novel prognostic marker for predicting cancer, sepsis, CVD, and post-operative complications
- NRBC is helpful to identify hemolytic anaemia, leukemoid reaction, sepsis, thalassemia
![f4 f4](https://erba.velocitabrandconsultants.com/wp-content/uploads/elementor/thumbs/f4-14-qoh6kj9tj97l6znc2jysss9lv8wktafka1tph0i2rk.png)
Diluting Mechanism
ADDM (Auto Dispense Diluting Mechanism)
Your Benefit:
Auto dispenses predefined volume, reducing contamination and manual errors
![f5 f5](https://erba.velocitabrandconsultants.com/wp-content/uploads/elementor/thumbs/f5-14-qoh6kj9tj97l6znc2jysss9lv8wktafka1tph0i2rk.png)
Reporting Formats
Customized reporting formats
Your Benefit:
Saves morphological findings & other test results in single report
![f6 f6](https://erba.velocitabrandconsultants.com/wp-content/uploads/elementor/thumbs/f6-11-qoh6kj9tj97l6znc2jysss9lv8wktafka1tph0i2rk.png)
Storage
Storage: 1,00,000 Results with Histograms
Your Benefit:
More data storage allows an easy access to old patient samples for monitoring the clinical progress
![f7 f7](https://erba.velocitabrandconsultants.com/wp-content/uploads/elementor/thumbs/f7-8-qoh6kk7nq38villyx2dfda12gmry0zjam6h6yagolc.png)
Linearity
Linearity : WBC - 0~300×10³/μL, PLT - 0 - 3000 x 10^9/L
Your Benefit:
Wide linearity in H360 helps to clinically analyze extreme patient demographics with leukemias, thrombocytosis etc.
Complete blood count analyser for mid-size laboratories and those looking to upgrade from 3 part differential system
Technical Specification
Mode Type
Whole blood, Predilute, Capillary
36 Parameters
CBC : WBC, RBC, Hb, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PCT, PDW-CV, PDW-SD, P-LCR, P-LCC.
DIFF : Neutrophils % & #, Lymphocytes % & #, Monocytes % & #, Eosinophils % & #, Basophils % & # *ALY % & #, *LIC % & #, *NRBC % & #, *CWBC, *NLR, RDWI & MENTZER INDEX,*Research use only
Principle of measurement
RBC/PLT: Electrical Impedance
WBC + DIFF: Tri-angle Laser Flowcytometry
Hb: Cyanide-Free Colorimetry
MCV: Measured
HCT: Calculated
Histograms
3 (WBC/RBC/PLT)
Scattergram
3 (2D ), 1 (WBC/Baso)
Throughput
upto 60 Tests/Hr
QR Code Reader
For reading sample bar codes & QC values
Sample Volume
Whole blood : 15 µL
Pre-diluted : 20 µL
Capillary : 20 µL
CBC Mode : 11 µL
Linearity Range
WBC : 0 - 300 ( x 103/µL)
RBC : 0.00 – 8.50 (x 103/µL)
Hb : 0 - 25.0 (g/dL)
HCT : 0 - 67 (%)
PLT : 0 - 3000 (x 103/µL)
Calibrator
ELite H Cal
Tri-level Controls
Elite H5 CON L, N, H
Open Vial Stability at 2-8˚C
16 Days
Data Storage
100,000 Results with Graphs, 100 calibrator results, Reagent log storage for 6 months
Quality Control
500 files with L-J graph
X-Bar Analysis
Dimension (mm)
364x498x431
Weight (Kg)
26.5
Interfaces
4 USB ports (For connecting external devices including external printer) + 1 LAN Port (HL 7, TCP/IP; LIS connectivity)
Operating Environment Temperature
15-40˚C
Humidity
30-90%
Power Requirement
A.C.100-240V with built-in voltage stabiliser
50/60Hz, 200VA
System General
System Type
Discrete, automated, random access, patient prioritized, clinical chemistry analyzer, 1,2, 3 reagent system
Throughput
Upto 1040 tests/hr (800 tests/hr) photometric and 240 tests/hr with ISE (4 Channel)*
Analytical Methods
1-Point, 2-Point, Rate-A, Rate-B, Direct Potentiometry
Barcode Reader
For reagents and samples
Reaction Tray
147 hard glass cuvettes
Reaction Mixing
Stirrer with variable speed
Reading Volume
Reaction volume 150 ul
On-board Laundry
7 stage cleaning, 2 stage drying
Photometer
Multi-wavelength diffraction grating with 15 wavelengths (340, 376, 405, 450, 480, 505, 528, 546, 570, 600, 628, 660, 700, 750 and 800 nm)
Calibration
K-Factor, Linear (1, 2 point & multipoint), 4P and SP Logit-log, cubic spline, exponential, polynomial. On-board serial dilution for calibrator
OD Range
0.0 - 3.0
Light Source
Tungsten halogen lamp
Detector
Silicon photo-diode
Water Consumption
<25 litres /hr
Programmable Parameters:
Default system pack parameters + upto 99 user defined parameters, Unlimited profiles and unlimited calculation items
Quality Control
QC plot data with QC rules. Provision for lab mean. Twin plot
Power Requirement
AC220V± 10%,50 Hz or AC 110 V± 10%, 60 Hz
Power Consumption
2000VA.
Dimensions (mm)
Approx . 1350 (L) x 1000 (D) x 1250 (H)
Weight
Approx . 275 kgs.
Sample Handling
Sample
Serum, Urine, Plasma, CSF, whole blood, other.
Standard / STAT Carousal
44 positions
Sample Rack
Upto 150 samples can be placed simultaneously (excluding standards, calibrators & controls)
Sample Pipetting
2 -60 µl (adjustable in 0.1 µIstep) for Biochemistry, 70 µl (fixed) for ISE Capacitance probe with level sensing, vertical obstruction detection, Clot detection and serum indices
Auto Rerun
Repeat with same, increased or decreased volume (upto 1:150)
Sample Tubes/ Cups
Primary tubes of 5 ml, 7 ml, 10 ml and sample cups
Reagent Handling
Reagent Tray
86 reagent positions with on-board refrigeration
Reagent Pipetting
Two separate probes for Rl, R2 & R3 Rl:60-300 µl (adjustable in 0.5 µl)
R2
0, or 10- 200 µl (adjustable in 0.5 µl)
R3
0, or 10- 200 µl (adjustable in 0.5 µl) Capacitance probe with liquid level sensing & vertical obstruction detection
System Interface
Analyzer - PC
RS - 232 C Serial Port PC - Host
Computer
Bidirectional TCP/IP & RS-232
PC - Printer
USB
Operating system
Windows-based
Database
Unlimited results